HomeTo Order
 

BOD AnalysisCalorimetersCOD AnalysisDissolution TestingNitrogen and Sulfer Analyzers

SpectroscopyTestingTitration and PHTOC Analyzers

9000 Series Nitrogen / Sulfur Analyzers
9000 Series Nitrogen/Sulfur Analyzers

Our 9000 Series Analyzers use proven technology to perform total nitrogen and/or total sulfur analyses from low ppb to percent levels. Their speed, accuracy, and precision are unrivalled — more stable than coulometric methods, more sensitive than x-ray, safer and faster than Kjeldahl.

Cat. No. AK-99821N – 115v, AK-99821S – 115v, AK-99821NS – 115v

Features:

  • One injection – two results
  • Nitrogen, sulfur or simultaneous nitrogen/sulfur
  • Fast, precise measurement of solid, liquid & gas samples
  • Sensitivity from low ppb to % levels
  • Superior to wet chemistry and other instrumental methods
  • No environmentally hazardous catalysts or reagent
  • Include Windows®-based operating & data capture software and computer system
  • Patented Pyro-chemiluminescent™ nitrogen and Pyro-fluorescent™ sulfur technology, compliant with many ASTM and DIN methods

Applications:

Typical applications include adipic acid, asphalts, biodiesel, coal, diesel, feces, corn products, gasoline & additives, LPG & natural gas, lube oil & additives, petroleum fractions, pharmaceuticals, polymers & additives, protein solutions, water/waste water, foods & beverages, pesticides, soil, starch, urine, wood & paper, and many more.

Specifications:

Detection Method
9000N Nitrogen Analyzer: Pyro-chemiluminescence™
9000S Sulfur Analyzer: Pyro-fl uorescence™
 
Analytical Range*
9000N Nitrogen Analyzer: ppb to 17%
9000S Sulfur Analyzer: ppb to 40%
 
Precision
±40 ppb or ±2% RSD, whichever is greater
 
Typical Sample Size*
Gas 10 cc
LPG 15 μl
Liquid: to 20 μl
Solid: up to 500 mg
*Analytical range and sample size may be a function of sample matrix and method of introduction
 
Typical Analysis Time
Gases / Liquids: less than one minute
Solids: 2 - 15 minutes (dependant upon sample type/weight)
 
Temperatures
Combustion zone: ambient to 1100°C in 1°C increments
 
Power
115 VAC 50/60 Hz 12 Amp
230 VAC 50/60 Hz 6 Amp
 
Gases
Dry oxygen: 300-600 cc/min 99.75%, 3 bar (40 psig), >5 ppm moisture
Dry carrier (argon or helium): 50-200 cc/min 99.99%, 3 bar (40 psig), > 5 ppm moisture
 
Dimensions & Weight
51.5 cm h (20.276”) x 47.0 cm w (18.504”) x 60.0 cm d (23.622”) 39.5 kg (87 lb)