TGA-50/50H and TGA-51/51H

Micro and Macro Series of Thermogravimetric Analyzers with Advanced Thermobalance Design

Features

  • Precise measurements to 1 µg
  • Efficient evolved gas analysis
  • Increased functionality: the TGA-51/51H analyzes up to 10 grams
    of sample while maintaining precision in the µg range
  • A wide selection of gas atmosphere from air, inert gas,
    reactive gas including hydrogen to vacuum
  • Extensive selection of sample crucibles

Specifications:

  TGA-50 TGA-50H TGA-51 TGA-51H
Temperature Range Ambient to 1000°C Ambient to 1500°C Ambient to 1000°C Ambient to 1500°C
Weight Measuring Range ±20mg, ±200mg ±20mg, ±200mg, ±2000mg
Readability 0.001mg
Sample Mass 1g including tare 10g including tare
Programmable Heating Rate 0.1°C/hour to
99.9°C/min
0.1°C/hour to
50.0°C/min
0.1°C/hour to
99.9°C/min
0.1°C/hour to
50.0°C/min
Hold Time 0 to 999min, 0 to 999 hour
Temperature program format 99 steps maximum
Temperature program file up to 100 files
Cooling Method Air-cooled
Signal Output Analogue and digital
Atmosphere Control Air, inert gas, reactive gas or vacuum. Built-in gas flowmeter (250ml/min maximum)
Dimensions W173xD550xH500mm W173xD600xH540mm
Power Supply AC100V, 120V, 220V, 240V, 500VA, 50/60Hz AC100V, 120V, 220V, 240V, 1.2kVA, 50/60Hz AC100V, 120V, 220V, 240V, 1kVA, 50/60Hz AC100V, 120V, 220V, 240V, 1.5kVA, 50/60Hz