Our 745 Nitrate/Nitrite Reducer utilizes vanadium (III) to rapidly reduce nitrite and nitrate to nitric oxide (NO). The resulting NO can then be detected and quantitated using ANTEK's
Model 7050 Nitric Oxide Analyzer or 9000 Series Nitrogen/Sulfur Analyzers.
Cat. No. AK-99503-115V/230V
Features:
-
Complete, rapid conversion
-
Wide dynamic range
-
Unaffected by dilution
-
Handles aqueous or hydrocarbon samples
-
Well suited for large samples
-
Built-in programmable heating block
-
Durable glassware
Covering a wide dynamic range and unaffected by dilution, the 745 is well suited for large sample sizes and easily handles aqueous or hydrocarbon samples. Durable glassware adds to its
simplicity and ease of operation.
Applications:
The 745 is effectively incorporated into analysis of biological fluids, chemical solutions, tissue bath solutions, wastewater, cell media, petrochemicals, tissue homogenates, and more.
Specifications:
|
Detector Response Time
|
40 milliseconds
|
|
Limits of Detection
|
Liquid: < 50 pg
|
|
Typical Sample Size
|
Liquid: 1 mL - 20 mL
|
|
Typical Reduction Time
|
Liquids: 0.5 - 3 minutes dependent upon concentration
|
|
Heater Temperature
|
Ambient (NO2-), 95°C (NO3-)
|
|
|
|
Requirements |
|
Power
|
115/230 VAC, 50/60 Hz, 3/2A
|
|
Gases
|
Dry helium: 1.5 bar (20 psi), Purity > 99.99%
|
|
|
|
Physical |
|
System Dimensions
|
height: 34.0 cm (13.5 in) width: 40.6 cm (16 in)
depth: 25.4 cm (10 in)
|
|
Weight
|
7.3 kg (16 lb)
|
|