HomeTo Order
 

BOD AnalysisCalorimetersCOD AnalysisDissolution TestingNitrogen and Sulfer AnalyzersSpectroscopyTestingTitration and PHTOC Analyzers

FTIR 8400S
Advantage 8400 | FTIR Spectrophotometer

Discover the Advantages of the FTIR 8400

Shimadzu was the first company to offer dynamic auto-alignment as a standard feature in an FTIR spectrometer in its' price range. With state-of-the-art technology, the Shimadzu dynamic alignment system continuously maintains optimal alignment during data acquisition, ensuring consistently reproducible spectra without the need for tedious mechanical adjustments.

Cat. No. SZ-220-93112-01

The Signal-to-Noise Advantage
Unprecedented S/N for an FTIR in its price-class; peak-to-peak S/N is 20,000:1 or better, guaranteed!

The Dynamic Alignment Advantage
Optimal interferometer alignment is maintained automatically for continuous optimization and exceptional reproducibility.

The Validation Advantage
The software validates instrument performance to ensure compliance. Shimadzu's unique GLP window permanently records all scan parameters and data manipulation events with each acquired spectrum.

The Simplicity Advantage
New FTIR operators will love the quick-start function that prompts the user through every action from setting scan parameters and acquiring the spectrum, to detecting peaks and printing the results.

The IR Solution Advantage
The powerful new IRsolution software increases the ease of use of the IRPrestige. The comprehensive package includes standard data processing functions, spectral search, quantitation and a easy report generation. Tailor the software to fit your application with the customizable user interface to suit a range of users.

Specifications:

Item FTIR-8400
Interferometer Michelson type with 30° incident angle, dynamic alignment, sealed and desiccated.
Optical System Single beam optics
Beam Splitter Germanium on KBr plate
Light Source High brightness ceramic
Detector Temperature-controlled high sensitivity detector (DLTAGS detector)
Signal-to-Noise Ratio Greater than 20,000:1
(KRS-5 plate, 4 cm -1 resolution, 1 minute scan, 2200 cm -1, P-P)
Wavenumber range 350 cm -1 ~ 7800 cm -1
Resolution 0.85,2,4,8,16 cm -1
Mirror speed 3 steps: 2.8 mm/sec, 5 mm/sec, 9 mm/sec.
Data sampling Using He-Ne laser
Sample compartment W 200 mm x D 230mm x H 170mm
Data processing functions Arithmetic operations, peak pick, spectrum subtraction, smoothing, baseline correction, data cut, data correction, normalization, Purity check, Kubelka-Munk conversion, Kramers-Kronig analysis, ATR correction, Fourier transform, derivatives, transmittance-absorbance conversion, peak area integration, peak ratio calculation, emission correction, deconvolution, Macro Program (BASIC), JCAMP conversion, ASCII conversion, logarithmic conversion, wavelength-wavenumber conversion, shift along X-axis
Optional programs Quantitation, multilinear regression quantitation, spectrum search, PLS Quantitation, Curve Fitting, Mapping Program
Validation programs Provided as standard Polystyrene auto-shutter for use in full automatic mode is available as an option
Ambient conditions Temperature: 15°C - 30°C
Humidity: Below 70%
Power requirements AC 100V/120V/220/230/240, 50/60Hz, 160 VA
  Privacy Policy | Comments