HomeTo Order
 

BOD AnalysisCalorimetersCOD AnalysisDissolution TestingNitrogen and Sulfer AnalyzersSpectroscopyTestingTitration and PHTOC Analyzers

UV-2450/2550
UV-2401/2501 | A Superbly Engineered Instrument

A Superbly Engineered Instrument

Features

This single monochromator system provides the analyst with the reliability and performance our customers have come to expect in Shimadzu instruments. The low stray light, wide dynamic range, small beam size combination makes this instrument well suited to handle a variety of practical situations.

Software

From small sample volumes to highly absorbing samples, the UV Probe software gives you the power and versatility of an extremely sophisticated system in a package that's easy to learn and use. Additional software enhancements like Kubelka-Munk conversion, area calculations and peak picks make this an instrument for the real world. What's more, it can resolve the triplet peaks in the vapor spectrum of benzene near 250nm with ease.


The UV-2550

An Unparalleled Instrument For Your Demanding Applications
When you can't dilute the sample or reduce the path length, think of the UV-2550. The double blazed grating, double monochromator (DDM) in this instrument is just waiting for those samples. Shimadzu has achieved the world's lowest level of stray light and highest energy throughput with this unique design. By creating two blaze wavelengths (200nm and 600nm), Shimadzu has overcome the energy loss and sensitivity restrictions usually associated with double monochromator instruments. The face of the grating is actually divided into two unique areas for short and long wavelength measurements. The sodium iodide spectra for the double monochromator UV-2550 and a single monochromator instrument described below put this information into practical terms.

Specifications

Wavelength Range

190 to 900nm (performance guaranteed range). Extendable to 1,100nm through the use of the optional photomultiplier. (The measurable range maybe restricted in the shorter wavelength side depending on the type of photomultipler used.)

Monochromator System

UV-2450: Single monochromator with a high-performance blazed holographic grating in the aberration corrected Czerny-Turner mounting.

UV-2550: Double monochromator with a high-performance double-blazed holographic grating in the aberration corrected Czerny-Turner mounting.

Resolution

0.1nm

Spectral Bandwidth

0.1, 0.2, 0.5, 1, 2 and 5nm

Wavelength Accuracy

±0.3nm

Wavelength Repeatability

±0.01nm

Wavelength Scanning Speed

FAST, MEDIUM, SLOW, and SUPER SLOW

Light Source

50W halogen lamp (2,000 hours of life) and D2lamp (500 hours of life)

Light Source lamp switching

Selectable between 282nm and 393nm

Stray Light

UV-2450: Less than 0.015% at 220nm and 340nm
UV-2550: Less than 0.0003% at 220nm and 340nm

Detector

Photomultiplier R-928

Photometric System

Double beam, direct ratio system with dynode feedback

Photometric Mode

Absorbance (Abs.), transmittance (%), reflectance (%) and energy (E).

Photometric Range

Abosrbance: -4~5 Abs. (0.001 Abs. increments)
Transmittance: 0~999.9% (0.01 increments)
Reflectance: 0~999.9% (0.01% increments)

Photometric Accuracy

±0.002Abs(0~0.5Abs), ±0.004Abs(0.5~1Abs),±0.3T (0~100%T)
(all determined with NIST 930D standard filter)

Photometric Repeatability

0.001Abs (0~0.5Abs), ±0.1%T

Baseline Correction

Selectable with storage in firmware

Baseline Flatness

Within ±0.001Abs.
(excluding noise, 2nm slit width and SLOW wavelength scanning speed)

Drift

Less than 0.0004 Abs. per hour (after 2 hours warm-up)

Dimensions

570 (W) x 660 (D) x 275 (H) (mm)

Weight

36kg

Power Supply

AC 100V/120V/220V/240V, 50/60Hz 250VA (swithc-selectable)

 

  Privacy Policy | Comments