TN-4110
TN-4110

On-line Water Quality Analyzer

TOC is used in a variety of applications from management of waste water treatment plant influent and effluent, to drinking water supply management and monitoring of impurities in process and natural waters. And the range of applications in which this type of on-line technology is used is expected to continue expanding.

In addition, because nitrogen is considered to be one of the factors which contributes to the over-proliferation of organisms in enclosed bodies of water, management of these public water sources and of treatment plant influent and effluent also includes demand for monitoring of nitrogen levels.

Features

  • Automatic dilution expands the analysis range and allows the reduction in salt, acid and alkali concentrations to help reduce maintenance and extend catalyst life
  • A complete line of options designated to accommodate samples with suspended particles, slime or seaweed reduces the occurrence of trouble and remedial maintenance in on-line TOC measurement.
  • Optional multi-stream selectors allow analysis of 2 streams, or up to 6 different streams, of varying carbon concentrations.
  • Measurement condition and measurement interval of each sample stream individual programmable
  • User defined calibration check or re-calibration. Automatic calibration is incorporated using standard solutions stored on-board.
  • On-board carrier/combustion gas purification allows use of air supplied from a compressor, thus eliminating expensive and potentially dangerous cylinder gases.

Applications

  • Process management for waste stream influent and effluent at treatment facilities.
  • Operation management of all types of plant waters (i.e.: cooling water, recirculating water, boiler water, wash/rinse water, condenser water)
  • Continuous monitoring of drinking/ process water, and natural water (rivers, lakes, etc.)
  • Continuous monitoring or organic contamination in water streams that must be reported to regulatory agencies.
Backwash Strainer Sample Pretreatment Unit Backwash Strainer Sample Pretreatment Unit
  • Simple and effective sampling unit for one, homogenous and low contaminated sample stream
  • Sample passes through the strainer into a sampler to supply to the analyzer
  • After sampling, air pushes the sample out of the sampler to prevent adhesion of contaminants and acts to back wash the strainer to prevent the strainer from clogging
   
Single Stream Suspended Solids Pretreatment Unit Single Stream Suspended Solids Pretreatment Unit
  • Process management for waste stream influent and effluent at treatment facilities.
  • Operation management of all types of plant waters (i.e.: cooling water, recirculating water, boiler water, wash/rinse water, condenser water)
  • Operation management of all types of plant waters (i.e.: cooling water, recirculating water, boiler water, wash/rinse water, condenser water)
  • Continuous monitoring of drinking/ process water, and natural water (rivers, lakes, etc.)
  • Continuous monitoring or organic contamination in water streams that must be reported to regulatory agencies.
   
Multi-Stream Suspended Solids Pretreatment Unit Multi-Stream Suspended Solids Pretreatment Unit
  • Sampling works similar as above described
  • Up to 6 sample stream switching is possible
  • Different types of measurement and different full scale ranges can be set for each sample stream
  • A wide range of sample concentrations may be measured on the multiple streams because of the unique rinsing system
  • The TN-4110 is high performance TN analyzer which uses the combustion/chemiluminescence detection technique
  • The TN measurement range is from 0-5ppm full scale to 0-200ppm full scale with the dilution function up to 4000ppm
  • With the short 4 minutes measurement cycle sudden changes in TN concentration can be caught
Specifications:
Measurement type Total Nitrogen TN
Measurement principle Combustion decomposition with chemiluminescence detection
Measurement range Variable from 0-1 ppm full scale to 0-200 ppm full scale (0-4000 ppm is possible with on-board dilution function)
Reproducibility Within +/- 4 % of full scale at less than 4 ppm full scale, within +/- 2 % full scale at greater then 4 ppm full scale
Span stability Within +/- 2 % of full scale/day (with temperature fluctuation within 5 °C)
Zero stability Within +/- 2 % of full scale/week
Measurement cycle Minimum 4 minutes
Sample dilution Dilution with syringe, variable sample dilution factor of 2-20
Auto calibration 1-2 point auto calibration possible
Multi-stream option 2 streams with optional sample switching set, Maximum 4 streams with optional flow line switcher
Display LCD 40 characters x 14 lines
Recorder option Recorder width is 100 mm, 6 dot recorder
Printer option Thermal type, 40 columns, chart width with 110 mm.
Analog output Selectable output of 0-1 VDC, 0-16 mA or 4-20 mA (insulated output)
Output RS 422
Alarm Selectable among upper/lower limit, upper/upper limit and lower/lower limit and instrument trouble alarm
Input signal Sample measurement start, sample measurement stop, calibration start, alarm reset
Event signals Ready state, operating online, measuring, calibrating and more.
Carrier gas Air or oxygen, supply pressure: 300 kPa, free of dust, oil mist, water droplets
Power 100,110,115,120,220,230,240 V as ordered; 5A ;50-60 Hz
Construction Indoor wall mounting (optional stand available)