SALD-201V

Cat. No. SZ-346-62325-11 SALD-201V Model 1
Cat. No. SZ-346-62325-51 SALD-201V Model 2

 

The SALD® 201V Particle Size Analyzer is a compact, economically
priced instrument that offers high performance and reliability features for
a wide range of food, liquid, cosmetic, and pharmaceutical applications.
A measuring range of 0.25 to 350µm accommodates most routine
analysis requirements.

 


Software

WingSALD software for Windows® provides flexibility for setting up test samples and enables real time display of particle size distribution. Enhanced data processing includes superimposed comparison of up to 12 graphs to establish trends and confirm reproducibility.

 

Specifications

Method of Measurement Laser diffraction method
Measuring Range 0.25 - 350µm
Optical System Light Source Semiconductor laser (670 nm wavelength)
Sample Cell Batch Cell Pyrex glass and tetrafluoroethylene
Flow-thru cell Fused silica, SUS304, and tetrafluoroethylene
Sampler Sample Bath SUS304, 200cm3 capacity
Liquid Pump Peristaltic pump
Material of
wetted parts
SUS304, silicone rubber, and polypropylene
Power Requirement Analyzer
main unit
100,110,120,200,220 or 240 VAC as ordered, 100VA
Sampler 100,110,120,200,220 or 240 VAC as ordered, 100VA
Dimensions (mm) Analyzer
main unit
280w x 340d x 510h 25kg
Sampler unit 240w x 330d x 290h 11kg
Environmental requirements Temperature 10-30°C
Humidity 20-80%
PC Hardware requirements IBM 100% compatible Windows® 95; CD-ROM drive; Pentium® 16MB RAM; 20MB of free HD space;
1 serial port (when the sampler is to be controlled by PC)
  Privacy Policy | Comments