Cou-Lo™ Trans  
The Cou-Lo <em>Trans</em> Coulometric Karl Fischer Titrator

The revolutionary Cou-Lo Trans is a portable, compact and very easy to operate coulometric Karl Fischer titrator.

Cat. No. GRS-21000

Features:

A revolution in water content determination

The Cou-Lo Trans is preprogrammed for the determination of water content in Transformer / insulating oil samples. Absolutely no data input is required and there is only one button for the operator to press. This simple 'one touch' operation makes the Cou-Lo Trans so easy to use that it requires no specialist knowledge or training to use it effectively.

Truly portable

Supplied as standard with a printer, low drift cell and ruggedised carry case, the Cou-Lo Trans has been specifically designed for outdoor use. Mobility is further enhanced by some flexible power options - the Cou-Lo Trans can be powered from the mains, by its own internal battery or by means of its car adapter.

The correct result, every time

At the heart of the Cou-Lo Trans is GR Scientific's state of the art A CE Control System . The patent pending ACE Control System (Automatically Compensated Errors) guarantees that the electrolysis current produced and the count rate displayed are always correctly synchronized, regardless of changes to the electrolysis cell resistance.

By eliminating the need to calibrate the reagent the ACE Control System also adds further to the Cou-Lo Trans' remarkable user friendliness.

Specifications:

Titration method coulometric Karl Fischer titration
Electrolysis control A CE Control System
End point detection AC polarisation
End point indication visual display
print out
acoustic beep
Display 40 character alphanumeric backlit LCD
Measuring range 1µg - 100mg water (possible)
1µg - 10mg water (typical)
Moisture range 1ppm - 100ppm water
Max. sensitivity 0.1µg
Max. titration speed 2 mg per minute
Max. current 400 ma
Drift compensation automatically controlled
Start delay time preset 10 seconds
Power supply 90 - 264 V AC, 47-63 Hz
12V DC car adapter
12V DC internal battery
Precision 10µg - 100µg ± 3µg
100µg - 1mg ± 5µg
above 1mg ± 0.5%
Calculation modes volume / density, preset values
Display format ppm (mg / Kg)
Print format µg, ppm (mg / Kg)
Method storage preset method
Printer 42 character high speed thermal printer
Stirrer speed microprocessor controlled
Battery Life 8 hours running time
Dimensions 250 x 245 x 120 mm
Weight 3 Kg (without carry case)
Carry case standard
Calendar/Clock analysis time and date print out
Battery Low display and print out indication