HomeTo Order
 

Gas Chromatography

Ion ChromatographyLiquid ChromatographyThin Layer Chromatography
   

GC-2014

High Performance and Reliability for Capillary and Packed Column Applications

Shimadzu's versatile GC-2014 gas chromatograph offers exceptional performance, value, and ease of use. Ideal for routine analysis, the GC-2014 utilizes the high-end technology of the GC-2010 to cover a wide range of applications from capillary to packed column analysis.

Features:

Superior Performance

Advanced Flow Control
The GC-2014 has the same proven superior Automatic Flow Control as the GC-2010, resulting in unsurpassed retention time reproducibility. This digital control is standard for both capillary and packed columns. Manual detector gas control is optional for most detectors on the GC-2014.

Detectors
Incorporates the GC-2010 detector designs for capillary analyses and the GC-14 detector designs for packed columns. The TCD unit is ideal for packed column measurements, employing the semi-diffusion cell design used in the TCD-14. The newly designed FPD-2014 takes advantage of the holophotal construction of the FPD-2010.

Column Oven
By using the GC-2010 control electronics and cooling mechanisms, column oven performance has been greatly improved. Oven capacity is increased while maintaining a small bench space, with enough capacity to accommodate both capillary and packed columns together.

Flexibility

Shimadzu's GC-2014 provides users with the flexibility to expand their systems for any application. The model for our System GC program, it offers an array of injection units and detectors, with the ability to install up to four detectors and three injectors.

Injection Units
With the GC-2014, the optimum injection mode can be selected from four types of injection units. Users can choose from among a dual-packed, split/splitless, single packed, or direct injection unit.

Detectors
Choose from a full line of detectors: FID, ECD, TCD, FTD (NPD), FPD, PID, PDID, ELCD, or SCD. The system is equipped with a detector control system optimized for dual detectors (FID and TCD) and comes standard with auto-ignition and re-lite features.

Mounted under the top cover for added stability, injectors and detectors are easily accessible from the top of the GC. A variety of option units, including the AOC-20 Series, help support a wide range of analyses.

Easy Operation

Large LCD Window
The LCD window shows graphically all of the operation of the GC, including the data results, temperature profile, and the chromatogram in progress.

Intelligent Self-Diagnostics
The GC-2014 provides complete self-diagnostics functionality and complete logging functions, which are compliant with GLP/GMP requirements.

Digital Control
Using the electronic flow controller, the column inlet pressure, column flow rate, linear velocity and split ratio can be digitally set. Flow meters have now become obsolete. Users can easily set carrier gas flow, and can choose between Advanced Pressure Control and low-cost manual flow controllers for the control of detector gas.

Specifications:

Column Oven

Temperature Range

(Ambient + 10°C) ~ 400°C (using liquid CO 2 gas: -50°C - 400°C)

Dimensions

250 (W) x 360 (H) x 175 (D) mm

Oven Capacity

15.8L

Temperature Accuracy

Set value (K) ±1% (calibration at 0.01°C increments)

Temperature Deviation

2°C max. (on 200mm dia. circumfrence 30 mm from rear)

Temperature Variation Coefficient

0.01°C/°C

Temperature Program Steps

Up to 20 (cooling program possible)

Programmed rate setting range

-250°C ~ 250°C/min

Total time for all steps

9999.99 minutes max.

Linear Heating Range

30°C/min up to 150°C
20°C/min up to 250°C
10°C/min up to 380°C
7°C/min up to 400°C (at 25°C ambient temperature)

Cooling Rate

300°C ~ 50°C in 6 min max. (at 25°C ambient temperature)

Columns Accepted

Capillary Columns: 2
Packed columns for GC14B: 4 (Glass columns:2)

Sample Injection Unit

Temperature Range

Up to 400°C

Heating Settings

1°C steps

No. of units installed simultaneously

Up to 3 units

Sample injection unit types

Dual packed, single packed, split/splitless

Carrier Gas Flow Controller

For Packed/Dual

Flow rate setting range

0 ~ 100mL/min

Programmable Steps

7

Programmed rate setting range

-400 ~ 400mL/min

Correction Function

Maintains column flow rate during column oven heating

For Capillary Split/Splitless
(Split/Splitless injection mode)

Pressure Setting Range

0 ~ 970kPa

Programmable Steps

7 (pressure-decreasing program possible)

Programmed rate setting range

-400 ~ 400kPa/min

Total flow rate setting range

0 ~ 200 mL/min

Correction Function

Maintains column average linear velocity during column oven heating (for capillary only)

(Pressure mode direct injection)

Pressure Setting Range

0 ~ 970kPa/min

Programmable Steps

7

Programmed rate setting range

-400 ~ 400kPa/min

(Flow-rate mode direct injection)

Flow rate setting range

0 ~ 1200mL/min

Programmable Steps

7

Programmed rate setting range

-400 ~ 400mL/min

Correction Function

Maintains column average linear velocity during column oven heating (for capillary only)

Detectors

Temperature Range

400°C max. (FID, TCD, FTD) 350°C max. (ECD, FPD)

Temperature Setting

1°C steps

No. of units installed simultaneously

Up to 4 units (restricted depending on detector type)

Detector Type

FID, TCD, ECD, FPD, FTD for capillary/packed

Flame Ionization Detector (FID)

System

Dual flow rate differential system

Temperature Range

400°C max.

Minimum detected quantity

3pgC/s (dodecane)

Dynamic Range

10 7

Nozzle

Quartz glass Standard: for packed, Option: for capillary

Thermal Conductivity Detector (TCD)

System

Dual flow rate differential system

Temperature Range

400°C max.

Dynamic Range

10 5

Sensitivity

40,000 mV - mL/mg (built-in pre-amplifier, with 10 x amplification) .

Electron Capture Detector (ECD)

System

Fixed current system using 63Ni370MBq radiation source

Temperature Range

400°C max.

Minimum detected quantity

0.1pg/s

Dynamic Range

10 4

Flame Photometric Detector (FPD)

Temperature Range

350°C max.

Dynamic Range

P: 10 4
S: 10 3

Minimum detected quantity

P: 0.5pgP/s (tributyl phosphate)
S: 8pgS/s (dodecane thiol)

Flame Thermionic Detector (FTD)
(Two types, one for capillary and one for packed. The specifications are the same.)

Temperature Range

400°C max.

Dynamic Range

N: 10 3
P: 10 3

Minimum detected quantity

N: 0.4pgN/s (azobenzene)
P: 8pgP/s (malathion)

Display

Dimensions

240 x 320 dot graphics display (30 characters x 16 lines)

Dimensions, Weight, Power Requirements (GC main unit)

Dimensions

400 (W) x 690 (H) x 607 (D) mm

Weight

48 kg (GC-2014AF model)

Power Requirements

AC100V/120V 230V
1800VA (GC-2014AF model) or
2600VA (GC-2014AF model), 50/60Hz

  Privacy Policy | Comments