Rugged, economical, fixed-wavelength detector for analytical or preparative HPLC and LC. A front panel knob allows selection of the wavelength to monitor the absorbency of almost any
sample.
|
Manufacturing Standards
|
Meets applicable safety and EMC certification standards; UL and CE certified
|
|
Autozero Range
|
–0.5 to + 1 AU, to within 5 x 10 -5 AU; front panel or remote activation
|
|
Communication Interface
|
Event and autozero inputs; three channels output data to a recorder
|
|
Detector Type
|
UV silicon photodiode
|
|
Drift
|
3 x 10 -4 AU/hr
|
|
Environmental Conditions
|
Indoor use Altitude: up to 2000 m Temperature range: 5–40°C Air pressure: 75–105 kPa Pollution degree: 1 or 2, in accordance with IEC 66 Humidity: Maximum relative
humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C
|
|
Event Marker
|
Negative deflection on trace; front panel or remote activation
|
|
Flow Cells
|
The following fused quartz flow cells are available:
|
Flow Cell
|
Path (mm)
|
Volume (µL)
|
psi
|
|
HPLC
|
10.0
|
11.0
|
1000
|
|
LC
|
10.0
|
40.0
|
1000
|
|
Microbore
|
5.0
|
1.3
|
1000
|
|
Preparative
|
2.0
|
10.0
|
1000
|
|
Preparative
|
0.5
|
2.5
|
1000
|
|
Preparative
|
0.1
|
0.3
|
1000
|
|
|
Flow Cell Sensitivity
|
<4 x 10 -4 AU for flow step of 0.5–2 mL/min with methanol
|
|
Front Panel
|
LCD indicator shows percent of full-scale output
|
|
Lamps
|
Standard: Mercury (254 nm), life of 2000 hours; Phosphor-coated Mercury (280 nm), life of 500 hours
|
|
Lamp Warm-up Time
|
One hour maximum
|
|
Monochromator
|
Dual beam with interference filter, to monitor at a single wavelength • Range: 254 and 280 nm standard (selectable on front panel) •Spectral bandwidth: 0.2 nm at 254 nm; 12
nm at 280 nm
|
|
Noise with Air Block Installed
|
254 nm, short term, peak-to-peak, 4 x 10 -5 AU/cm; long term, peak-to-peak, 6 x 10 -5 AU 280 nm, short term, peak-to-peak, 6 x 10 -5 AU/cm; long term, peak-to-peak, 6 x 10 -5 AU
|
|
Power Requirements
|
Frequency: 50–60 Hz Voltage: 100–120 or 220–240V; mains voltage fluctuations not to exceed ±10% of the nominal voltage Current rating: 1.0A for
100–120V, or 0.5A for 220–240V
|
|
Reference Chamber
|
Air
|
|
Sensitivity Range
|
0.001 to 1.0 AU with standard flow cell installed. Up to 100 AU when 0.1 mm flow cell is used. Selectable on front panel.
|
|
Temperature Sensitivity
|
3.0 x 10 -4/°C for temperature change from 21° to 24°C
|
|
Time Constant
|
0.1, 0.2, 0.5, 1, 2, or 5 seconds; selectable on front panel
|
Dimensions
(w x d x h)
|
32 x 32 x 13 cm (12.6 x 12.6 x 5.2 in.)
|
|
Weight
|
8 kg (17.6 lbs.)
|