Gilson's 15X Series of UV/VIS Detectors deliver versatile, rugged detection for your HPLC. The optical design lets you optimize the detection of your sample at any wavelength from 190 to
700 nm.
Model 151: GA-10105311 – Programmable variable wavelength UV/Vis detector with single wavelength detection: includes analytical cell and built-in display
Model 155: GA-10105511 – Programmable variable wavelength UV/Vis detector with single, dual and scan wavelength detection; includes analytical cell and built-in display
Model 152: GA-10105411 - Programmable variable wavelength UV/Vis detector with single wavelength detection: includes analytical cell. Controlled by Unipoint or TRILUTION Software (sold
separately)
Model 156: GA-10105611 – Programmable variable wavelength UV/Vis detector with single, dual and scan wavelength detection; includes analytical cell. Controlled by Unipoint or
TRILUTION software (sold separately)
Features:
-
Interchangeable flow cell design accommodates capillary to preparative HPLC applications
-
The detectors have a sensitivity range of 0.001 to 2.0 AUFS
(Absorbance Units, Full Scale)
-
Controlled via Gilson’s UniPoint™ System Software or choose stand-alone keypad control with the 151 and 155 models
-
Stores up to 10 detection methods
-
Choose from the following modes to specify detection conditions and basic operating parameters:
-
single wavelength
-
file
-
status
-
setup
-
dual wavelength (155 and 156)
-
scan wavelength (155 and 156)
Specifications:
|
Manufacturing Standards
|
Meets applicable safety and EMC certification standards; UL and CE certified
|
|
Autozero Range
|
Suppresses up to 1.0 AU with 5 mm flow cell installed
|
|
Communication Interface
|
Event and autozero inputs; remote contact closure can be activated by other instruments or the computer. Three channels output data to a recorder.
|
|
Drift
|
After 1 hour at constant temperature: 3.0 x 10-4 AU/hr
|
|
Environmental Conditions
|
Indoor use
Altitude: up to 2000 m
Temperature range: 5–40°C
Air pressure: 75–105 kPa
Pollution degree: 1 or 2, in accordance with IEC 66
Humidity: Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C
|
|
Flow Cell Assemblies
|
The following quartz flow cells are available:
|
Flow Cell
|
Path (mm)
|
Volume (µL)
|
psi
|
|
Analytical
|
5
|
12
|
500
|
|
Microbore
|
2
|
1.6
|
500
|
|
Preparative
|
0.2
|
0.7
|
500
|
|
Preparative
|
0.05`
|
0.16
|
500
|
|
|
Flow Sensitivity
|
4.0 x 10-4 AU for flow step of 2.0 mL/min to 0.5 mL/min with methanol
|
|
Front Panel
|
Four-line display shows modes, parameters, commands, and data. Six soft keys and HELP, ESC, CLEAR, ENTER, numeric, and arrow hard keys (151/155 only). LED indicator lights for POWER,
UV, VIS , REMOTE, and ERROR. LAMP ON/OFF hard key.
|
|
Lamps
|
UV lamp: Deuterium, warranted for 750 hours or 180 days
Visible lamp: Tungsten/halogen, warranted for 500 hours
|
|
Lamp Warmup Time
|
One hour maximum
|
|
Linearity
|
1%, when value is within sensitivity range of 0.001 to 2.0 AU
|
|
Monochromator
|
Dual beam, stepper motor-driven
-
Range: 190–700 nm
-
Spectral bandwidth: 9 nm
-
Setting accuracy: ±2 nm
-
Setting precision: ±0.2 nm
|
|
Noise with Air Block Installed
|
220 nm, short term, peak to peak, 2.0 x 10-5 AU/cm; 254 nm, short term, peak to peak, 2.0 x 10-5 AU/cm; 350 nm, short term, peak to peak, 10.0 x 10-5 AU/cm; 415 nm, short term, peak
to peak, 8.0 x 10-5 AU/cm; 520 nm, short term, peak to peak, 6.0 x 10-5 AU/cm; 650 nm, short term, peak to peak, 4.0 x 10-5 AU/cm
|
|
Peak Width
|
0 and 4 to 99 seconds for single- or scan-wavelength mode; 4 to 99 seconds for dual-wavelength mode
|
|
Power Requirements
|
Frequency: 50–60 Hz
Voltage: 90–120 or 220–240V; mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 1.0A for 90–120V, or 0.5A for 220–240V
Online switching regulation of line input from 80 to 260V AC
|
|
Sensitivity Range
|
0.001 to 2.0 AU. Sensitivity is adjustable in increments of 0.001 AU
|
|
Software Control
|
Via GSIOC or RS-232. You can issue commands to do the following:
-
Select all detection parameters
-
Program parameter changes during run to optimize detection
-
Send digital detector output
|
|
Static RI Sensitivity
|
Methanol versus cyclohexane at 270 nm: 5.0 x 10-3 AU
|
|
Temperature Sensitivity
|
3.0 x 10-4/°C for temperature change from 21° to 24°C
|
Dimensions
(w x d x h)
|
26.5 x 43.5 x 15.6 cm (10.4 x 17.1 x 6.2 in.)
|
|
Weight
|
8.6 kg (19 lbs.)
|
|