The 849 Multiple Injection Module is designed for parallel injection onto an HPLC or LC/MS system. Deliver high-throughput capacity with outstanding speed and performance. Modules are the ideal choice for high-throughput injection onto FIA/MS.
Manufacturing Standards |
Meets applicable safety and EMC certification standards; UL and CE certified |
Available Sample Loops |
5, 10, 20, 50, 100, 200 or 500 µL; 1, 2 or 5 mL |
Available Valves |
Rheodyne RV700-100 or Rheodyne RV700-112 |
Carryover † |
RV700-100 Valve
• <1 ppb with the following rinse volumes:
– injection port rinse volume=250 µL
– rinse station rinse volume=500 µL
RV700-112 Valve
• <0.05% with the following rinse volumes:
– injection port rinse volume=1000 µL
– rinse station volume=1000 µL |
Dead Volume |
Injection Port |
Dead Volume* (µL) |
0.7 mm OD probe tip |
11 |
1.3 mm OD probe tip |
15 |
1.5 mm OD probe tip |
14 |
* From probe to valve (including calibrated tubing)
|
Environmental Conditions |
|
Front Panel |
Power light, 4 or 8 INJECT lights and 4 or 8 LOAD lights |
Liquid-Contact Materials |
304 and 316L stainless steel, alumina ceramic, FEP, PEEK |
Power Requirements |
Frequency: 50–60 Hz
Voltage: 90–240V; mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 2.0A for 90–120V or 1.0A for 220–240V (849); 3.0A for 90–120V or 1.5A or 220–240V (889) |
Precision † |
RV700-100 Valve
• Partial loop fill: CV <2.0% with a 500-µL syringe and 10-µL partial loop fill in a 20-µL injection loop
• Total loop fill: CV <5.0%
RV700-112 Valve
• Partial loop fill: CV <2.0% with a 5-mL syringe and 1-mL partial loop fill in a 2-mL injection loop
• Total loop fill: CV <5.0% |
Communication Interface |
GSIOC |
Valve Switching Speed |
RV700-100: <135 mS
RV700-112: <370 mS |
Dimensions (w x d x h) |
61 x 18 x 13 cm (24 x 7 x 5 in.) |
Weight (with installed valves) |
7.7 kg (17 lbs.) with four installed valves
11.4 kg (25 lbs.) with eight installed valves |
† Contact Mandel service to learn what method was used to obtain the value.