HomeTo Order
 

Gas ChromatographyIon ChromatographyLiquid Chromatography

Thin Layer ChromatographyGel Permeation Chromatography
   
215 Direct-Inject Liquid Handler/Injector
Image of 215 Direct-Inject Liquid Handler/Injector

The 215 Direct-Inject Liquid Handler/Injector is designed to handle the demands of smaller injection volumes—delivers microliter volumes with precision and accuracy. Capable of injecting volumes of 500 nL to 5 mL from 384- or 96-well microplates, vials or tubes.

 

 



Features:

  • Accommodates up to 17 standard- or deep-well microplates
  • Increases throughput with valve switching speeds five times faster than the standard 215 Liquid Handler/Injector
  • Optional Peltier Racks for cooling and heating of samples in standard microplates
  • Ideal for injecting 1 µL (or less) directly into the 841 Injection Module to minimize carryover and maximize precision; can also accommodate large-volume injections without reconfiguring the instrument
  • Add a second 841 Injection Module to increase throughput while preserving bench space

Specifications:

Manufacturing Standards Meets applicable safety and EMC certification standards; UL and CE certified
Arm Speed X: >45.7 cm/sec (>18 in./sec)
Y: >40.6 cm/sec (>16 in./sec)
Communication Interface RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs, and one switched +24V DC 1A output
Liquid Handling Reproducibility:
1.37% CV with 50 µL of water
Accuracy:
99.7% with 50 µL of water
Front Panel 8-character display, START and emergency STOP soft keys
Horizontal Motion Strength X: 5 kg (11.1 lbs.)
Y: 7 kg (15.6 lbs.)
Injection Carryover <0.05% with 100-µL syringe and total loop filling method
Injection Reproducibility 1.5% CV
Injection Valve Switching Speed <100 ms
Injection Volume 500 nL–5 mL
Locator Plate Capacity Up to five Code 200-Series racks, up to seven Code 20- or 30-Series racks (or a combination of up to five racks of both types), or one Code 500-Series and one Code 200-Series rack
Maximum Syringe Flow Rates for Water
Syringe Size (µL) Max. Flow Rate (mL/min)
100 3.37
250 8.43
500 16.87
1000 33.75
5000 100.00
10,000 100.00
25,000 100.00
Power Requirements Frequency: 50–60 Hz
Voltage: 100–120 and 200–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 2.0A for 100–120V or 1.0A for 220–240V
Probe Positioning Performance Accuracy: ±0.5 mm in X/Y dimensions, ±1 mm in Z dimension
Repeatability: ±0.25 mm in X/Y/Z dimensions
Probe Rinse Through a dedicated rinse station for rinsing the inside and outside of the probe; selectable rinse volume and flow rate. Optional inserts for level-sensing, nonleveling-sensing and flow rinse
Sampler Type X/Y/Z with stationary rack design
Software Via Gilson UniPoint™ System Software or TRILUTION™ Software
Syringe Pump Integral, high-precision, single-piston
Syringe Capacity 100, 250 or 500 µL; 1 or 5 mL
System Resolution 0.1 µL
Vertical Punch Strength 4.1 kg (11.0 lbs.)
Dimensions 91.4 x 61 x 55.8 cm* (38.5 x 24 x 22* in.)
*Maximum height. Z-arm height is adjustable to accommodate vessel heights between 1 and 150 mm (dependent on installed Z-arm).
Weight 39.9 kg (89 lbs.)

† Contact Mandel Customer Service to learn what methods and conditions were used to obtain the values.