HomeTo Order
 

Gas ChromatographyIon ChromatographyLiquid Chromatography

Thin Layer ChromatographyGel Permeation Chromatography
   

350 Micro Pump

Gilson’s 350 Micro Pump provides precise splitless solvent delivery for nano to micro LC that delivers accurate and reproducible gradients from 300 nL/min to 50 µL/min The 350 is capable of delivering true nano flow rates without requiring a flow splitter.

Cat. No. GF-38103931

Features:

  • Modular, binary pumping system with independent micro mixer optimizes separation with nano, capillary and micro columns for 1-D, 2-D and LC/MS/MS applications
  • Splitless flow delivery eliminates the inaccuracy of flow splitters, while dramatically reducing solvent consumption
  • Dual-piston pump with patented magnetic coupling ensures precise, accurate flow rates from 300 nL/min to 50 µL/min
  • 60-nL mixer for fast gradient response times
  • Pump head seal wash permits the use of buffered solvents with increased confidence and decreased maintenance

Applications:

  • Nano- and micro-scale HPLC and LC/MS/MS analysis
  • Two-dimensional HPLC separations for biological and drug-related assays
  • MALDI-TOF plate preparation via HPLC

Specifications:

Pump Options

Isocratic pump (350D): electrically-actuated valves, 300 nL/min to 50 µL/min, 30 MPa (300 bars, 4,350 psi), micro degasser
Gradient pumping system (350D + 350 + gradient pumping kit): electrically-actuated valves, 300 nL/min to 50 µL/min, 30 MPa (300 bars, 4,350 psi)

Pump Heads

Micro E head

Number of Solvents/Pump

One

Flow Rate

Gradient: 0.3–50 µL/min (4,350 psi)
Isocratic: 0.15–50 µL/min (4,350 psi)
Selectable units: nL/min or µL/min
Adjustment increment: 0.01 µL/min
Accuracy: <±1% (5–50 µL/min)
Repeatability: <0.5% for 5–50 µL/min; <1% for 0.5–5 µL/min

Composition and Mixing

Adjustment increment: 0.1% B for 1–50 µL/min; 1% B below 1 µL/min
Dwell volume/Effective mixing volume: 60 nL (up to 5 µL/min); 3 µL (above 5 µL/min)
Gradient linearity range: 2% at 1% B; 97% at 100% B
Accuracy: ±1%
Repeatability: <1% RSD

Pressure

Operating range: 0 MPa–30 MPa (300 bar, 4,350 psi)
Accuracy: <±1% or 0.1 MPa (1 bar, 15 psi)
Repeatability: <1% or 0.1 MPa (1 bar, 15 psi)

Pump Volume

37.8 µL (from inlet of inlet valve to outlet of outlet valve)

Piston Stroke Volume

39.4 µL

Software

UniPoint™ System Software via 506C interface

Communication Interface

Nine-pin GSIOC socket, three input and four output contact closures

Front Panel

Green POWER, yellow ON, and yellow REMOTE LED indicators; ON/OFF soft key

Liquid Contact Materials

316L stainless steel, titanium, hastelloy, zirconium, FEP, PTFE, PCTFE, ETFE, PEEK, and UHMWPE

Power Requirements

Frequency: 50–60 Hz
Voltage: 100–120 or 220–240V; mains voltage fluctuations not to exceed ±10% of the nominal voltage

Environmental Operating Temperature

4–40°C

Manufacturing Standards

Meets applicable Safety and EMC certification standards; UL and CE certified

Dimensions (w x d x h)

26 x 34 x 30 cm (10 x 13 x 12 in.)

Instrument Weight

350: 13.5 kg (30 lbs.)
350D: 15 kg (33 lbs.)

Shipping Weight

350: 18.5 kg (41 lbs.)
350D: 20 kg (44 lbs.)

  Privacy Policy | Comments