Gilson’s
flexible liquid handler and injector is engineered to
improve throughput for drug discovery and drug metabolism
laboratories. High-throughput capacity is ideal for
parallel injection onto HPLC or LC/MS systems.
Manufacturing Standards |
Meets applicable safety and EMC certification standards; UL and CE certified |
Arm Speed |
X: >45.7 cm/sec (>18 in./sec)
Y: >40.6 cm/sec (>16 in./sec) |
Communication Interface |
RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs and one switched +24V DC 1A output |
Injection Volume |
5 µL–5 mL |
Injection Reproducibility |
<2.0% CV with 500-µL syringe and partial loop filling method; <5.0% CV with 500-µL syringe and total loop filling method |
Liquid Handling |
Reproducibility:
0.57% CV with 500 µL of water
Accuracy:
99.2% with 500 µL of water |
Front Panel |
8-character display, START and emergency STOP soft keys |
Horizontal Motion Strength |
X: 5 kg (11.1 lbs.)
Y: 7 kg (15.6 lbs.) |
Injection Carryover† |
<0.05% |
Locator Plate Capacity |
Up to five Code 200-Series racks or one Code 505 or 505H rack plus one Code 200-Series rack |
Maximum Syringe Flow Rates for Water
(with eight syringes installed) |
Syringe Size
(µL) |
Max. Flow Rate
(mL/min) |
100 |
0.42 |
250 |
0.80 |
500 |
2.11 |
1000 |
4.22 |
2500 |
8.40 |
5000 |
12.50 |
|
Maximum Syringe Flow Rates for Water
(with four syringes installed) |
Syringe Size
(µL) |
Max. Flow Rate
(mL/min) |
100 |
0.75 |
250 |
1.50 |
500 |
4.25 |
1000 |
8.50 |
2500 |
16.75 |
5000 |
25.00 |
|
Power Requirements |
Frequency: 50–60 Hz
Voltage: 100–120 and 200–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 2.0A for 100–120V or 1.0A for 220–240V |
Probe Positioning Performance |
Accuracy: ±0.5 mm in X/Y dimensions, ±1 mm in Z dimension
Repeatability: ±0.25 mm in X/Y/Z dimensions |
Probe Rinse |
Through a dedicated rinse station for rinsing the inside and outside of the probes |
Sampler Type |
X/Y/Z with stationary rack design |
Software |
Via Gilson UniPoint™ System Software or TRILUTION™ Software |
Syringe Capacity |
100, 250 or 500 µL; 1, 2.5 or 5 mL |
Syringe Pump |
Integral high-precision, eight-piston syringe pump |
Vertical Punch Strength |
0.6 kg/probe (1.4 lbs./probe) |
Dimensions (w x d x h) |
97.8 x 61 x 55.8* cm (38.5 x 24 x 22* in.)
*Maximum height. Z-arm height is adjustable to accommodate vessel heights between 1 and 150 mm (dependent on installed Z-arm). |
Weight |
39.9 kg (89 lbs.) |