HomeTo Order
 

PipettesPipettorsPipette TipsAccessoriesAutomated Liquid Handlers

InjectorsFraction CollectorsPumps & Tubing

   
Quad-Z 215 Liquid Handler/Injector

Based on the same robust design as the 215 Liquid Handler/Injector, the Quad-Z 215 features four independent probes with variable horizontal spacing from 9 to 18 mm for access to a wide range of sample vessels.

Cat. No. GA-25101412


Features:

  • Large bed capacity holds up to 17 microplates
  • Accommodates many of Gilson's racks for maximum flexibility for sample transfer between tubes, vials and plates
  • The 444 QuadDilutor has four independently controlled syringe pumps for use with the four independently controlled probes on the Quad-Z 215 for maximum flexibility across all four channels
  • The 444 accepts syringe sizes ranging from 100 µL to 25 mL
  • Optional 849 Multiple-Injection Module enables automated on-line injection onto parallel HPLC systems
  • Optional Vacuum Rack is ideal for SPE applications

Applications:

  • MALDI spotting
  • Reformatting
  • Cherry picking
  • Peptide synthesis

Specifications:

Manufacturing Standards

Meets applicable safety and EMC certification standards; UL and CE certified

Liquid Handling

Reproducibility:
2.1% CV with 1 µL of DMSO
Accuracy:
104% with 1 µL of DMSO

Injection

Volume:
5 µL–5 mL
Reproducibility:
2.0% CV with 500-µL syringe and partial loop filling method; <5.0% CV with 500-µL syringe and total loop filling method
Carryover:
<0.05%

Communication Interface

RS-232 or GSIOC; Four inputs (contact closure, TTL, or open-collector), four relay outputs and one switched +24V DC 1A output

Arm Speed

X: >45.7 cm/sec (>18 in./sec)
Y: >40.6 cm/sec (>16 in./sec)

Front Panel

8-character display, START and emergency STOP soft keys

Horizontal Motion Strength

X: 5 kg (11.1 lbs.)
Y: 7 kg (15.6 lbs.)

Locator Plate Capacity

Up to five Code 200-Series racks in the standard locator plate, up to five combined Code 200-series and Code 20/30-Series racks in the standard locator plate, up to seven Code 20- or 30-Series racks in the Code 20- or 30-Series locator plate, or one Code 505 or 505H rack plus one Code 200-Series rack

Power Requirements

Frequency: 50–60 Hz
Voltage: 100–120 or 220–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 2.5A for 100–120V or 1.4A for 220–240V

Probe Positioning Performance

Accuracy:±0.5 mm in X/Y dimensions±1 mm in Z dimension
Repeatability:±0.25 mm in X/Y/Z dimensions

Sampler Type

X/Y/Z with stationary rack design

Software

Via Gilson TRILUTION™ Software

Syringe Capacity

100, 250 or 500 µL; 1, 5, 10 or 25 mL

Syringe Pump

External, integral, high-precision, four-piston (444 QuadDilutor)

Vertical Punch Strength

3.06 kg/probe (6.74 lbs./probe)

Dimensions (w x d x h)

Quad-Z 215: 91.4 x 61 x 60.8* cm (36 x 24 x 24* in.)
*Maximum height. Z-arm height is adjustable to accommodate vessel heights between 1 and 150 mm (dependent on installed Z-arm).
444 QuadDilutor: 40.4 x 28.7 x 31.2 cm (15.9 x 11.3 x 12.3 in.)

Weight

Quad-Z 215: 40 kg (89 lbs.)
444 QuadDilutor: 14 kg (31 lbs.)

  Privacy Policy | Comments