Mandel Hopmepage
HomeTo Order
 

PipettesPipettorsPipette TipsAccessoriesAutomated Liquid HandlersInjectors

Fraction CollectorsPumps & Tubing

   

819 Injection Module

The 819 Injection Module enables you to inject samples onto Gilson's 215 Liquid Handler. The module contains a high-pressure, two-position switching valve that can also be used for fraction collection. It is an effective way to collect fractions when the flow rate is higher than 20 mL/min.

Cat. No. GA-251511

Features:

  • Carryover is less than 0.01%
  • Ideal for analytical to preparative injections
  • Make-before-break rotor seal design prevents pressure spikes from injection when operating at high flow rates

Specifications:

Valve Switching Speed

<500 ms

Injection Reproducibility*

<0.9% CV with 1-mL syringe and total loop filling method

Injection Carryover*

<0.01%

Available Valve

Rheodyne 7010

Available Sample Loops

5, 10, 20, 50, 100 and 200 µL; 1, 2 and 5 mL

Front Panel

LOAD and INJECT LED indicator lights

Communication Interface

RS-232 or GSIOC; one contact closure output and one contact closure input

Power Requirements

Frequency: 47–63 Hz
Voltage: 100–120 or 220–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Current rating: 1.0A for 100–120V or 5.0A for 220-240V

Liquid-Contact Materials

316L and 304 stainless steel, FEP, PEEK

Software

Via Gilson UniPoint™ System Software or 735 Sampler Software

Dimensions (w x d x h)

19 x 20 x 10 cm (7.5 x 8 x 4 in.)

Weight (with installed valve)

4 kg (9 lbs.)

 

 

*Contact Mandel service to learn what methods were used to obtain the values.

  Privacy Policy | Comments
Mandel Contact Us Mandel Homepage