The 845Z/845ZPREP Injection Modules give you the best of both worlds—high-throughput capacity with minimal carryover. Attaches directly to the Z-arm of the 215 Liquid Handler for aspirating sample directly into the loop.
Manufacturing Standards |
Meets applicable safety and EMC certification standards; UL and CE certified |
Available Sample Loops |
5, 10, 20, 50 and 100 µL |
Available Valve |
Rheodyne RV700-100 |
Communication Interface |
GSIOC; One contact closure input and one contact closure output |
Dead Volume * |
~40 µL from probe tip to valve |
Injection Volume |
100 µL–10 mL |
Front Panel |
One-character LED display and POWER light |
Injection Carryover * |
0.01% |
Injection Reproducibility * |
Total loop fill: CV = 0.5% |
Liquid-Contact Materials |
Stainless steel, PEEK, Teflon® |
Power Requirements |
+24V DC at 2.0A supplied via a 2.1-mm ID power plug; center contact is positive |
Software |
Via Gilson UniPoint™ System Software version 3.2 or higher |
Valve Switching Speed |
845Z: <266 mS
845ZPREP: 250 mS |
Dimensions (w x d x h) |
Control Box (installed on 215 control cabinet): 15.2 x 13.7 x 7.1 cm (6 x 5.4 x 2.8 in.)
Valve Box (installed on the 215 Z-arm): 7.9 x 19.3 x 11 cm (3.1 x 7.6 x 4.3 in.) |
Weight (with installed valve) |
2 kg (4.4 lbs.) |
*Contact Mandel service to learn what methods were used to obtain the values.