Model (Reference #) |
Volume (µL) |
Gilson Maximum Permissible Error |
ISO-8655 Permissible Error |
Systematic error (µL) |
Random error (µL) |
Systematic error (µL) |
Random error (µL) |
F2 ( GF-F123770) |
2 |
±0.08 |
≤0.03 |
±0.08 |
≤0.04 |
F5 ( GF-F123771) |
5 |
±0.1 |
≤0.04 |
±0.125 |
≤0.075 |
F10 ( GF-F123772) |
10 |
±0.1 |
≤0.05 |
±0.12 |
≤0.08 |
F20 ( GF-F123604) |
20 |
±0.2 |
≤0.06 |
±0.2 |
≤0.1 |
F25 ( GF-F123775) |
25 |
±0.25 |
≤0.07 |
±0.5 |
≤0.2 |
F50 ( GF-F123778) |
50 |
±0.4 |
≤0.15 |
±0.5 |
≤0.2 |
F100 ( GF-F123784) |
100 |
±0.8 |
≤0.25 |
±0.8 |
≤0.3 |
F200 ( GF-F123605) |
200 |
±1.6 |
≤0.3 |
±1.6 |
≤0.6 |
F250 ( GF-F123787) |
250 |
±3.00 |
≤0.75 |
±4.0 |
≤1.5 |
F300 ( GF-F123788) |
300 |
±3.5 |
≤0.75 |
±4.00 |
≤1.5 |
F400 ( GF-F123789) |
400 |
±3.6 |
≤0.8 |
±4.0 |
≤1.5 |
F500 ( GF-F123790) |
500 |
±4.0 |
≤1.00 |
±4.00 |
≤1.5 |
F1000 ( GF-F123606) |
1000 |
±8.00 |
≤1.3 |
±8.00 |
≤3.00 |
|