Model V15.17
Model V15.17

Designed for the separation of analytical and semi-preparative quantites of proteins in a single 17 x 15 cm polyacrylamide (PAGE) gels

Cat. No. BM-21080015

Features:

  • Runs a single gel
  • Built-in integral clamps hold the gel in place
  • Fits gels of different thickness with straight or notched glass plates

The V15.17 is a system for high-quality separation of proteins (and nucleic acids) in a vertical standard-gel format. The gel electrophoresis apparatus is supplemented by a complete selection of accessories and replacement parts, including an external gel casting system

The Model V15.17 is designed to electrophorese a single 17 x 15 cm (W x L) vertical gel for separation of up to 20 samples on a 0.8 mm; 1.5 mm or 3 mm thick gel.

The built-in integral clamps allow easy assembly and hold gels of different thickness leak-free in place. The V15.17 is the system of choice for laboratories with low sample throughput but the need of long running distances or large sample volumes (up to 4,000 µl).

It is recommended to use Gel Casting Clamps for simple, leak-free casting of gels. This allows casting of single gels in a fraction of the time of the "conventional" tape or grease and binder clip methods.

Specifications:

Apparatus dimensions
(W x D x H)
Gel size (W x L) 17 x 15 cm
Glass plate size (W x L) 19.7 x 16.0 / 18.5 cm
Number of gels 1
Capacity max. 20 samples
Typical sample volumes 34 – 4000 µl per well (depending on gel thickness)
Combs
Thickness (mm) Number of Wells Well Width (mm) Volume per Well (µl) b
0.8 1 prep. + 2 marker lanes 10 12 <14 20 139.7 13.0 10.0 <8.2 4.8 1000 c 94 72 <59 34
1.5 1 prep. + 2 marker lanes 10 12 <14 20 139.7 13.0 10.0 <8.2 4.8 2000 c 175 135 <110 63
3.0 1 prep. + 2 marker lanes 10 12 14 20 139.7 13.0 10.0 8.2 4.8 4000 c 350 270 220 126
Buffer volume 400 - 500 ml (upper buffer chamber) + 400 - 500 ml (lower buffer chamber)

b Volume calculated for an insertion depth of 9 mm
c Volume of the central preparative well